Device classification
Materials
Service life
Mode of operation
Ingress protection
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables.
Only use this Philips Avent single/double electric breast pump Advanced in combination with the adapter
supplied.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic
immunity
The Philips Avent single/double electric breast pump Advanced is intended for use in the electromagnetic
environment specified below. The customer or the user of the breast pump should assure that it is used in
such an environment. The essential performance functions are the accuracy of the vacuum pressure and the
backflow protection to withstand milk leakage into the motor unit. The Philips Avent single/double electric
breast pump Advanced was tested for immunity to electromagnetic disturbances and passed using the
criteria of no change in modes, vacuum accuracy (±20 %), backflow protection and operation of the breast
pump.
Single electric breast pump
Applicable classifications:
Degree of protection against electric shock;
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
ABS, silicone (motor unit)
Polypropylene, PBT, silicone (other parts)
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Double electric breast pump
Electromagnetic environment –
guidance
The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
English