M18™ 500GB CONTROL HUB
3.5.2 SYSTEM UPDATE - M18™ 500GB CONTROL HUB
Before updating a system, go to the ABOUT section of the SETTINGS menu and make a
note of the current firmware revisions in the system.
Firmware updates can be performed with the m18™ WIRELESS MONITOR or tablet
running the Milwaukee® Pipeline Inspection App.
1. Turn on both the M18™ 500GB CONTROL HUB and M18™ WIRELESS MONITOR or
tablet running the Milwaukee® Pipeline Inspection App.
2. Check that the WIRELESS MONITOR or tablet running the Milwaukee® Pipeline
Inspection App is connected to the CONTROL HUB by WI-FI. If it is not connected,
you will need to connect it.
3. Insert a Thumb Drive with the firmware update on it into the CONTROL HUB's USB-A
port.
4. Turn the navigation dial to SETTINGS
5. Turn the navigation dial to SYSTEM UPDATE, press the navigation dial to enter the
submenu.
SCREEN SETTINGS
DATE & TIME
UNITS
Feet
LANGUAGE
English
SYSTEM UPDATE
FACTORY RESET
ABOUT
SERVICE & SUPPORT
6. Select CONTROL HUB and press the navigation dial. The filename of the update
should be seen in this window. Press the navigation dial to begin the update.
SETTINGS
CONTROL HUB
WIRELESS MONITOR
Select CONTROL HUB
, press the navigation dial.
SETTINGS
MANAGE JOBS
2019/05/10 6 FILES
53
SETTINGS
MTCam_hub.bin
The firmware update is shown