Hazardous area Zone 2, 21 or 22
ambient temperature: -20°C to +60°C
11)
Load cell #1
Rmin = 300W
11)
Load cell #2
Rmin = 300W
Load cell #8
Rmin = 300W
Hazardous area Zone 0, 1, 20 or 21
ambient temperature: -20°C to +60°C
10)
Load cell
#1
10)
Load cell
#2
Load cell
Datum
Name
Date
Erstellt
14.11.18
Schallhorn
Written by
Geprüft
15.11.18
Hiller
Reviewed by
Freigabe
15.11.18
Schallhorn
Released by
Output parameters:
The output voltage is
given by the connected
evaluation unit.
11)
Equipotential bonding conductor
10)
#8
Equipotential bonding conductor
Material
.
Ausgabe / Revision
Änderung / Alteration
01
PA50181104
Cable junction box
6)
R1
6)
R2
Input parameters:
Um = Umax = 25V
For load cells in
6)
zone 2: input
R8
circuit in type of
protection Ex nA
Ex nA IIC T6 Gc
Ex tb IIIC T70°C Db
-20°C ≤ Ta ≤ 60°C
Ex tb IIIC T50°C Db
-20°C ≤ Ta ≤ 40°C
9)
Cable junction box
6)
R1
6)
R2
6)
R8
See page 2
9)
Benennung / Title
Safety Instructions
Zeichnungs-Nr. / Drawing number
36931-751-16
11)
Evalaution
unit
with
non
intrinsically
output
safe
to the load
cells
10)
Evalaution
unit
with
intrinsically
output
safe
to the load
cells
Maßstab /
Scale
1:1
Blatt
1
Sheet
von
3
of