3.1 General rules
E135
E136
E137
E155
E157
E158
E160
E161
E166
E172
E173
E174
E175
E176
E177
E178
E180
E181
E182
E183
E196
78
Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
Same buffer in CAL. Measured value of the second buffer differs less than
6 mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
XXX Bytes are missing so that the method or the silo line could not be stored
or not enough RAM for running TIP.
Exit: Send new command.
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
No sequence defined in TIP.
Exit. The error message disappears on next start.
A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
No new temporary variable.
Exit: The error message disappears on next start.
Measurement range of the secondary measured value (temperature)
exceeded. The primary measured value (pH, U, Ipol, Upol) can also be
unstable.
Exit: Rectify error or &m $S.
Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
In TIP a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
The warning interval of the internal buret D0 called.
Exit: Execute prep &a.b.p $G or start next method.
The warning interval of the external buret D1 called.
Exit: Execute prep &a.b.p $G or start next method.
The warning interval of the external buret D2 called.
Exit: Execute prep &a.b.p $G or start next method.
The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was
interrupted manually.
Exit. The error message disappears on next start.
Accessing to the memory card, the card was not (properly) inserted.
Exit: The error message disappears on next start.
The date of changing the battery of the card is expired.
Exit: The error message disappears on next start.
Memory card write-protected.
Exit: Send new command.
Memory card not formatted.
Exit: Send new command.
Memory card not accessible.
Exit: Send new command.
A directory with the same name exists already on the memory card.
Exit: Send new command.
Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
758 KFD Titrino