3.1 General rules

3.1.3 Status messages

Go:
$G
Hold:
$H
Continue:
$C
Ready:
$R
Stop:
$S
Detailed status conditions
Status conditions of the global $G:
$G .Mode.KFT .Inac:
.Req .Id1:
.Start:
.KFT1:
.Cond.Ok:
.Cond.Prog:
$G .Mode.SET...
$G .Mode.MEAS .Inac:
.Req .Id1:
.Meas:
$G .Mode.CAL .Inac:
.Req.Temp:
.Meas.Temp:
.Req.Buf1:
.Meas.Buf1:
.Req.Buf2:
.Meas.Buf2:
etc.
$G .Assembly.Bur
$G .Prep.X.Active:
.Empty.X.Active:
In TIP, its global status as well as the step number (X) is available.
$G .TIP.X
.Inac:
74
In order to have an efficient control by an external control device, it must also
be possible to query status conditions; they provide information on the status
of the Titrino. The trigger $D initiates output of the status. Status messages
consist of the global status, the detailed status and eventual error messages,
e.g. $S.Mode.SET;E26. The global status informs on the activity of the proc-
ess, while the detailed status conditions show the exact activity within the
process.
The following global status conditions are possible:
The Titrino is executing the last command.
The Titrino has been held ($H, key <meas/hold> or by an error which effects
the hold status).
The Titrino has been restarted actively after hold.
The Titrino has executed the last command and is ready.
A process has been aborted in an "unnatural manner". e.g. stopped or aborted
because there was an error.
Instrument at the beginning or at the end of a titration.
Instrument in the KFT mode, requesting Id1 after start.
Instrument in the KFT mode, requesting Id2 after start.
.Id2:
Instrument in the KFT mode, requesting Id3 after start.
.Id3:
Instrument in the KFT mode, requesting sample size after start.
.Smpl:
Instrument in the KFT mode, requesting unit of sample size after start.
.Unit:
Instrument in the KFT mode, processing the start conditions.
Instrument in the KFT mode, titrating to the first endpoint.
Instrument in the KFT, conditioning, endpoint reached (after the first
startup from the standby mode).
Instrument in the KFT mode, conditioning, endpoint not reached
(Conditioning progressing).
As KFT.
Instrument at the beginning or at the end of a titration.
Instrument in the MEAS mode, requesting Id1 after start.
Instrument in the MEAS mode, requesting Id2 after start.
.Id2:
Instrument in the MEAS mode, requesting Id3 after start.
.Id3:
Instrument in the MEAS mode, requesting sample size after start.
.Smpl:
Instrument in the MEAS mode, requesting unit of sample size after start.
.Unit:
Instrument in the MEAS mode, measuring.
Instrument at the beginning or at the end of a calibration.
Instrument in the CAL mode, requesting calibration temperature.
Instrument in the CAL mode, measuring calibration temperature.
Instrument in the CAL mode, requesting pH of buffer 1.
Instrument in the CAL mode, measures buffer 1.
Instrument in the CAL mode, requesting pH of buffer 2.
Instrument in the CAL mode, measures buffer 2.
Buret in filling process
.Fill:
Buret in DIS mode
.ModeDis:
Preparing buret X, X=0, 1, 2.
Emptying buret X, X=1, 2.
Instrument at the beginning or at the end of a TIP.