Metrohm 758 KFD Titrino Mode D'emploi page 107

Titreur pour la détermination de la teneur en eau et les titrages à point final rapides et précis avec grand écran lcd
Masquer les pouces Voir aussi pour 758 KFD Titrino:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

758 KFD Titrino
3.2.2.22. Mode.Parameter.TitrPara.Temp
Parameters for SET, KFT: Titration temperature in °C. If a Pt100 or Pt1000 is
connected, the temperature is measured continuously and the parameter
.Temp is updated.
The temperature is used for the temperature correction in pH measurements.
3.2.2.23. Mode.Parameter.TitrPara.TDelta
Parameter for SET, KFT: Time interval in s for the entry of a measurement
point in the list of measured points.
3.2.2.24. Mode.Parameter.StopCond.VStop.Type
Mode.Parameter.StopCond.VStop.V
Mode.Parameter.StopCond.VStop.Factor
Parameters for SET, KFT: Stop volume.
If an absolute stop volume (abs.) has been selected, the volume in mL is
valid.
A relative stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size ∗ factor
The factor is valid.
OFF means that the criterion is not monitored.
3.2.2.25. Mode.Parameter.StopCond.FillRate
Parameters for SET, KFT: Filling rate in the titration in mL/min. Max. means
maximum possible filling rate with the Exchange Unit in current use.
3.2.2.26. Mode.Parameter.Statistics.Status
Mode.Parameter.Statistics.MeanN
Mode.Parameter.Statistics.ResTab.Selected
Mode.Parameter.Statistics.ResTab.DelN
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN:
Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2 Remote control commands
–170.0...25.0...500.0
1...2...999999
abs., rel., OFF
0...99.99...9999.99
0...±999999
0.01...150, max.
original, delete n,
ON, OFF
2...20
delete all
1...20
103