Télécharger Imprimer la page

VOLTCRAFT VC-5080 Notice D'emploi page 36

Publicité

Les langues disponibles

Les langues disponibles

c) Resistance measurement
Make sure that the circuit components, circuits and components to be measured or any other
measuring objects are voltage-free and discharged.
Proceed as follows to measure resistance:
- Turn the rotary switch to the measuring range Ω. The small current values beneath
the measuring ranges indicate the measuring current.
- Connect the black test lead to the COM socket (6) and the red test lead to the VΩ
socket (7).
- Zero the multimeter using the adjustment screw (3).
- Make sure that the "Vrms/Vp-p" button (9) is not pressed.
- Connect both test leads and wait until the scale indicator has stabilized. The
display should show a value of approx. 0 ohm. Zero the meter to 0 ohm using the
0 ohm adjustment knob (8) in case of deviation. Always zero the multimeter when
changing measuring ranges. If a zero adjustment cannot be conducted, the mignon
(AA) batteries are in need of replacement.
- Read the measured value from the "Ω" scale. Multiply the displayed value by the measuring range to obtain the
correct reading (e.g. 100Ω (display) x10kΩ (measuring range) = 100 x 10,000 = 1MΩ (measured value).
- After measurement, turn the rotary switch to position "OFF" to turn off the multimeter.
36

Publicité

loading

Ce manuel est également adapté pour:

1218859