NOTE: If the actions listed below do not resolve the error, call the OneTouch
ERROR
POTENTIAL PROBLEM
The meter detected
a problem with the
colour of the test
strip.
All manuals and user guides at all-guides.com
12/3/03
1:27 PM
WHAT TO DO
1.
Repeat the test with a new test strip and:
◆
Apply blood to the pink test square only.
◆
Make sure the confirmation dot turns completely blue.
◆
Insert the test strip within 2 minutes of applying
blood.
Note:
If you continue to get any error message, com-
pare the colour of the test strip confirmation dot to the
Colour Chart on the test strip vial. If the glucose level is
very low (2.8 mmol/L) or very high (19.4 mmol/L),
follow the advice of your Healthcare Professional.
◆
Clean the meter and repeat the test with a new test
strip. See Caring for Your Meter for further informa-
tion.
2.
If these actions do not resolve the error, call the
®
OneTouch
49
Page 49
®
Customer Care Line at 1 800 663-5521.