Fuse replacement

Warning!
m
Patient hazard, equipment malfunction
and damage –
The replacement of fuses must be
referred to specialists as defined in
IEC 60364 or other applicable standards
(e.g. biomedical technicians, electricians,
electronics installers).
Fuses used must be
S# < 20,000: T1A L250Vac
S# >20,000: T2A H250Vac.
Before replacing fuses, turn off the ARTRO-
MOT®-SP3 and disconnect the device from
the power line.
Use an appropriate tool to remove the fuse
holder situated between the power switch
and the power connector (Fig. 1). Replace
the fuses and reinsert the fuse holder (Fig.
2). Ensure that the fuse holder properly locks
into place.
Fig. 1
Fig. 2
66