Scan list (for continuous mode only)
You are required to set up the scan list to include all desired analog input
channels. By default, the U2300A series scans only CH 101 with the following
settings.
The settings in channel configuration entry remain unchanged when the desired
data is sampled. You do not need to reconfigure the channel configuration entry if
you wish to sample new data using the same order and settings. The maximum
number of entries you can set is 100. Table below shows the structure of a scan list.
Table 3-2
Structure of a scan list with four entries
CHANNEL
108
101
103
102
To build a scan list
To build a scan list, follow the steps below:
– Use the ROUTe:SCAN command to define the list of channels in the scan list.
– Use the ROUTe:SCAN command if you wish to overwrite the initial setting of the
– To initiate a scan sequence, use either the DIGitize or RUN command.
To stop a scan initiated by the RUN command, use the STOP command.
Keysight U2300A Series User's Guide
– Range: ±10 V
– Input signal type: Single-ended
– Polarity: Bipolar
RANGE
10
±5
±10
±2.5
To determine what channels are currently in the scan list, use the
ROUTe:SCAN? query command.
scan list.
POLARITY
UNIP
BIP
BIP
BIP
Features and Functions
SIGNAL TYPE
SING
NRS
NRS
DIFF
3
45