Features and Functions
Sampling rate
Specify the sampling rate of each AI channel. Since the U2300A series DAQ
devices comes with multiplexing analog input, the maximum sampling rate
depends on the ADC's sampling rate and the entry number in the scan list.
For example, if four channels are specified in the scan list of the U2356A, the
maximum sampling rate is actually 500 kSa/s divided by four, which is 125 kSa/s
for each channel. However, in the U2331A, the maximum sampling rate is only up
to 1 MSa/s when switching of multiple channels is enabled.
Sample points
Specify the number of acquisition points for the channel. For example, if 800
sample points and four channels are specified in the scan list, there will be total of
3200 samples to be acquired.
The maximum sample points for single-shot acquisition are 8 MSa and for
NOTE
continuous acquisition is 4 MSa, where both types of acquisitions are of
continuous input mode.
44
Keysight U2300A Series User's Guide