Electromagnetic Compatibility Information
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601test level
± 8 kV
contact
± 2 kV,±4 kV,
± 8 kV,
± 15 kV air
± 2 kV ,
100kHz, for
AC power port
±0.5kV, ±1kV
(differential mode)
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°
, 270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
30 A/m ; 50 Hz or 60Hz
Compliance level
± 8 kV
contact
± 2 kV,±4 kV,
± 8 kV,
± 15 kV air
± 2 kV ,
100kHz, for
AC power port
±0.5kV, ±1kV
(differential mode)
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°
, 270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
30 A/m ; 50 Hz or 60Hz
EN
Electromagnetic envi-
ronment-guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power qua-
lity should be that of a
typical commercial or
hospital environment.
Mains power qua-
lity should be that of a
typical commercial or
hospital environment.
Mains power qua-
lity should be that of a
typical commercial or
hospital environment.
Power frequency
magnetic fields should
be at levels charactertic
of a typical location in
a typical commercial or
hospital environment.