TESTO 926 Mode D'emploi page 31

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

YES was selected:
- Automatic probe detection is started. Auto flashes
while the instrument looks for a radio probe that is
switched on.
- Once a radio probe is found, the probe ID is
displayed. If no probe is found, NONE lights up.
Possible reasons why probes are not found:
· The radio probe is not switched on or the
battery of the radio probe is spent.
· The radio probe is outside the range of the
measuring instrument.
· Sources of interference are influencing the radio
transmission (e.g. reinforced concrete, metal
objects, walls or other barriers between
transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
If necessary, rectify the possible causes for the
disruption to the radio transmission and start
automatic probe detection again with
- If further wireless probes are within reception range,
the probe ID of a different wireless probe may be
displayed.
If necessary: switch off other wireless probes or
remove from the reception range, and start
automatic probe detection again with
Press
2
4
To s s et A A uto O O ff:
Configuration mode is opened, Auto Off is lit.
Select the desired option with
:
· On: The measuring instrument switches off
automatically if no button is pressed for 10 min.
Exception: A recorded reading is shown on the
display (Hold or Auto Hold is lit).
· OFF: The measuring instrument does not switch
itself off automatically.
to change to the next function.
5. Operation
.
.
and confirm with
31