Motor type 21, 40,000 rpm
Motor type 21, 40,000 rpm/Min. speed
Motor type 21, 40,000 rpm/Max. speed 5000 – 40,000 rpm
Motor type 21, 40,000 rpm /Start ramp
Motor type 21, 40,000 rpm /Stop ramp
The maximum torque of the motor is boosted at the starting procedure of the reverse rotation to disengage stuck
implants or screws.
Reverse torque
Reverse torque/Increase
Reverse torque/Increase time
Resetting to
factory default
Default value/Set default value
Attention: Upon resetting to factory default all parameters will appear with factory default values
(except date and time).
Exit from Configuration Menu:
• To leave the Configuration Menu press "HP" and "PR" for 3 sec., until you hear the long tone.
18
Range
Factory
default
300 – 5000 rpm
300 rpm
40,000 rpm Set max. speed of the motor.
1 – 1000 ms/10,000 rpm 100 ms
1 – 1000 ms/10,000 rpm 50 ms
Range
5 – 30%
100 – 2000 ms
Choice
Yes/No
Definition
Set the min. speed of the motor.
Set acceleration time to 10'000 rpm
Set breaking time from 10'000 – 0 rpm
Factory
Definition
default
25%
Increase of the selected torque in reverse
rotation.
500 ms
Time during which torque is raised.
Factory
Definition
default
–
Resetting all parameters of the Configuration
Menu to factory default.
MD 11 Operating Instructions, Art. No. 31665, V10/14
MD 11