Before being used, thoroughly check the product. The same operation shall be carried out after cleaning.
-Check that the connection between the head and the handle is perfect and that the rheostat correctly
works.
-If the light is intermittent or in case it does not turn on, check the bulb, the batteries and the electric
contacts.
In case the bulb stops working, it is possible to replace it with original spare parts (cod. 31449).
To replace the bulb, it is necessary to take the speculum out, counter-clockwise turn the bulb and screw
the new one down (see picture). It is advisable to use only original bulbs.
Before removing the bulb, make sure that the instrument has been off for some minutes;
otherwise you run the risk of being burnt
-Check that the glass is clean; on the contrary it shall be removed. To do this counter-clockwise rotate
it and at the same time pull it towards you.
After cleaning, reintroduce the glass into its seat making the groove match with the small pivot and
clockwise rotating it (see picture).
-It is advisable to always have new bulbs and batteries at disposal for the maximum light.
-The otoscope head and handle can be cleaned with a cloth soaked with disinfectant, however they
cannot be dipped into liquids.
-The speculum can be sterilized in boiling water or in autoclave, in this case the temperature shall not
be over 121×; alternatively cold disinfectants can be used with the same concentration used to sterilize
medical instruments.
The product has been conceived to last forever; it is resistant to most part of chemicals except for
methyl ethyl ketones (M.E.K.), diluents and nail polish removers. All internal parts are brass and
stainless steel plated preventing rusting. If the lamp is exposed to humidity remove the batteries,
thoroughly dry to avoid short-circuits.
Thank you for having acquired our product.
This product meets the most stringent requirements regarding the selection of quality manufacturing
materials and also the final control. The product has a 12-month warranty, valid from the date it is
delivered by GIMA.
During the guarantee period free repair and/or replacement of any defective parts due to faulty
manufacture will be given, labour, postal, transport and packaging charges etc. are not included. The
guarantee therefore excludes components subject to wear and tear such as parts in rubber or PVC, etc.
No compensation will be given for loss of use of the product.
Furthermore replacement or repairs effected during the guarantee period do not lengthen the duration
of the guarantee.
This guarantee is void in the event of: repairs effected by unauthorised persons or with
spare parts not approved by GIMA, damage or defects caused by negligence, blows,
abnormal use of the appliance or faulty installation.
The guarantee will be invalidated if the registration number has been removed, cancelled or altered.
Faulty appliances must be returned only to the retailer where the item was bought. Any item dispatched
directly to us will be refused.