Programming Artromot ® -S3 Standard; Treatment Value Details - Standard Model - Ormed ARTROMOT-S3 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 84
5.2 Programming
ARTROMOT
Standard
Different programming levels are
provided to program the
ARTROMOT
®
-S3 Standard models.
You change between levels by pressing
the MENU key repeatedly.
The display always indicates the
currently selected level.
The following treatment values,
settings and information can be
entered/viewed on the programming
unit (1):
LEVEL 1:
- Abduction
- Adduction
- Internal rotation
- External rotation
LEVEL 2:
- Pause
- Therapy timer
- Speed
- New patient
LEVEL 3:
- Reverse on load - motor A
- Reverse on load - motor B
- Motor A ON/OFF
- Motor B ON/OFF
LEVEL 4:
- Transport setting
- Synchronized/
non-synchronized mode
- Total therapy time
- Service menu
®
-S3
MENU
MENU
MENU
60
Note!
• While you adjust the values, the
carriage will move to the set range.
This allows you to easily and
quickly determine the ROM where
the patient does not experience
pain.
• Special functions can be
programmed and retrieved with
ARTROMOT
®
-S3 Comfort devices
(see sections 5.4 and 5.5).
• The angle entered last for the
respective direction of motion will
be saved in each case.
5.3 Treatment value details –
Standard models
• To access the programming level you
need, press the MENU key repeat-
edly.
• The treatment parameters are selected
with the different parameter keys.
• To change the treatment values, press
the + / – keys.
• To enable/disable a function, press
the respective parameter key again.
• To save the entered values, press the
STOP key.
LEVEL 1:
I Abduction
Maximum value:
I Adduction
Maximum value:
Caution!
Patient hazard 
When using the shoulder restraint, be
sure not to program abduction values
greater than 80°.
175 degrees
30 degrees

Publicité

Table des Matières
loading

Produits Connexes pour Ormed ARTROMOT-S3

Table des Matières