5. Setting the treatment values
Warning!
Patient hazard
Before therapy, a test run consisting
of several exercise cycles must be
completed without the patient. Then
repeat the test run with the patient and
check that the movement does not
cause any pain.
Note: See also 2.2 and 2.3 as well as
page 249!
Note!
Programming is only possible when
the patient's chip card has been
inserted.
For information about therapy para-
meters and programming of special
functions, please refer to sections 5.1
to 5.5.
See section 5.6 for programming
examples.
Important!
The programming unit
ARTROMOT
®
-S3 "Graphics" can be
connected to all products of the
ARTROMOT
®
shoulder series with
serial numbers 3000 and higher.
It is possible to exchange chipcards
between the "Text" and "Graphics"
versions.
Please note:
1. If you use a "Graphics" program-
ming unit with a formatted chipcard
of the "Text" version, please note
the following
- the selected language of the"Text"
version remains activated"; it is
irrelevant for the "Graphics"
version
- the reverse on load setting saved
in the "Text" version is automati-
cally used for both motors
- the key lock is disabled.
2. If you use a "Text" programming
unit with a formatted chipcard of
the "Graphics" version, please note
the following
- the selected language automati-
cally changes to the default
setting: German
- the reverse on load setting saved
for motor A in the "Graphics"
version is automatically used for
both motors
- a key lock set in the "Graphics"
version will remain active,
because it is irrelevant for the
"Text" version.
5.1 General information on
programming
ARTROMOT
1. You activate the programming mode
by briefly pressing the MENU key on
the programming unit.
2. The treatment parameters and
functions are allocated to four
(Standard model) or six (Comfort
model) programming levels (four per
level). To be able to program a
parameter you will have to access
the corresponding programming
level. This is also done with the
MENU key. With each key press you
advance one level. The code M1, M2,
etc. that appears in the middle of the
display indicates the programming
level.
3. You activate the treatment parame-
ters and functions with the four
parameter keys below the display.
The symbols above the four parame-
ter keys indicate the assigned
parameters and functions.
58
®