Intermittent lighting up of the
LED, duration approx. 1 sec:
Slight glow:
2.5 [Fuego basic RF] Foot pedal programming
The laboratory burner and foot pedal can be newly programmed to each other if the
signal is interrupted or in cases of parallel use of several radio systems.
For programming of the radio channel, place the laboratory gas burner close to the
wireless foot pedal (max. 30 centimetres). Programming has a restricted range to
avoid inadvertent re-programming of other burners. (In cases where undesired
re-programming of a neighbouring burner occurs, switch it off during the
programming procedure.)
Programming procedure:
Adhere to the sequence for the programming procedure. Close the gas
!
regulator (2) by turning it clockwise to avoid inadvertent activation of the
flame.
1. Switch off the laboratory gas burner.
2. Switch on foot pedal by operating briefly or press again briefly.
3. Switch on the laboratory gas burner.
4. Shortly afterwards, press the programming key (20) on the foot pedal for approx.
1 second to start the programming procedure. The function LED (20) will now start to
glow. This procedure can last for between 0.5 and 30 seconds. As soon as the
laboratory burner and the foot pedal have been successfully programmed, the glowing
is switched off and the function LED (20) switches to brief, faint flashes.
The laboratory burner and the foot pedal have already been programmed to each
i
other by the manufacturers.
2.6 Switch-off the unit
The unit can be turned off by pushing the function knob (1) for more than 2 seconds.
[Fuego basic RF] The radio foot pedal is deactivated automatically.
To depressurize gas hose, shut off the gas supply and activate the burner
i
again until the flame extinguishes in order to burn the residual gas.
Afterwards turn off the gas burner at the function knob (1).
3. Application programs
This chapter explains the various application programs
("Button Start-Stop", "Pedal Standard" und "Pedal Start-Stop").
3.1 Button Start-Stop
The flame is ignited by operation of the function knob (1). The flame is extinguished
after renewed actuation of the function knob (1). Additionally the flame is automatically
extinguished when the burning timer has expired after 60 min.
28
Batteries are low (see paragraph 9).
Programming in progress (see paragraph 2.5).