Control Panel

Power
Button
Power On –
Green Light
System Warning
Description
Color
Power Button
Power Off –
None
Power On –
Green
System
Yellow
Warning
System Fault
Red
8
– Yellow Light
System Fault –
Red Light
Function
When the switch is turned so
that the "I" symbol is down,
the device is operational.
When lit, this indicates
the device has detected a
possible fault. The audible
alarm will not shut off until
the user stops the filling
process and checks the
device.
When lit, this indicates the
device has detected a fault
that requires service by the
provider. The device has
stopped filling and will not
begin filling again until the
problem is corrected.
Start/Stop Fill
Button
Fill Level
Indicators
Modes
Off – The device is turned off; mains power is
not connected.
Green light on – the device is on with no
faults detected.
Off – no system issue detected.
Flashing yellow with periodic audible beep
– the device has detected a low or high
pressure fault.
Steady yellow – the device has detected a
possible O
purity issue.
2
Steady yellow and flashing red with a
constant audible alarm – the device has
detected an O
purity issue with the oxygen
2
cylinder.
Off – No system faults have been detected.
Steady On with Constant Audible Alarm –
The device has detected a problem that
prevents it from operating. Contact your
home care provider for repair.
Cylinder