Some functional testers or patient simulators can be used to verify
whether the device works normally, for example, INDEX-2LFE Simulator
(software version: 3.00), please refer to the manual for the detailed
operation steps.
Some functional testers or patient simulators can measure the accuracy
of the device copied calibration curve, but they can not be used to
evaluate the device accuracy.
When using the device, please keep it away from the equipment which
can generate strong electric field or strong magnetic field. Using the
device in an inappropriate environment may cause interference to the
surrounding radio equipment or affect its working.
The measured accuracy will be affected by the interference of
electrosurgical equipment.
When several products are used on the same patient simultaneously,
danger may occur which is arisen from the overlap of leakage current.
CO poisoning will appear excessive estimation, so it is not recommended
to use the device.
This device is not intended for treatment.
The intended operator of the device may be a patient.
Avoid maintaining the device during using.
1. Overview
The pulse oxygen saturation is the percentage of HbO
blood, so-called the O
parameter for the respiratoy and circulatory system. A number of diseases
related to respiratory system may cause the decrease of SpO2 in the blood,
furthermore, some other causes such as the malfunction of human body's
self-adjustment, damages during surgery, and the injuries caused by some
medical checkup would also lead to the difficulty of oxygen supply in human
body, and the corresponding symptoms would appear as a consequence,
such as vertigo, impotence, vomit etc. Serious symptoms might bring danger
to human's life. Therefore, prompt information of patients' SpO2 is of great
help for the doctor to discover the potential danger, and is of great
importance in the clinical medical field.
Insert the finger when measuring, the device will directly display the SpO
value measured, it has a higher accuracy and repeatability.
concentration in the blood. It is an important bio-
2
in the total Hb in the
2