Sterilizing - Dentsply Sirona Midwest E Plus Notice D'utilisation

Masquer les pouces Voir aussi pour Midwest E Plus:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Sirona Dental Systems GmbH
Operating Instructions Midwest E Plus
66 04 206 D3686
D3686.201.01.01.02 08.2016
All manuals and user guides at all-guides.com
7.5

Sterilizing

Intervals:
Interval _ US
● Prior to initial operation
● Prior to every other use
Process
The instrument is clean and disinfected.
Action steps_instruments
Instrument maintenance is complete.
The instrument can be sterilized in packaging suitable for sterilization
and storage: paper/plastic composite packaging according to ISO
11607.
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Unit pressure in bar and psi_US
Temperature:
Overpressure:
Article
Wrapped instruments
Unpackaged
instruments
Gravity displacement steam sterilizers are permitted.
without DAC PROFESSIONAL _ US
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
After sterilizing
After sterilizing - heading
1. Remove the instrument from the steam sterilizer immediately.
Subsequent steps_instruments
CAUTION! The instrument is hot. Risk of burns!
NOTICE! Do not attempt to accelerate the cooling process by
immersing the instrument in cold water. This can damage your
instrument.
2. Store all instruments so that they are protected from contamination.
3. Sterilize again once the storage period has elapsed.
135° C (275° F)
2.13 bar (30.89 psi)
Holding time at 135 °C
Drying time
(275° F)
10 minutes
30 minutes
3.5 minutes
0 - 1 minute
7 Conditioning
7.5 Sterilizing