Télécharger Imprimer la page

A&D BA-T Serie Guide De Démarrage Rapide page 15

Balance analytique

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
8.
Typical Errors (Error Codes)
Display
Error code
Error 1
EC,E11
Error 2
Error 3
Error 6
EC,E16
Error 7
EC,E17
Error 8
Error 9
Con Err
Door Err
S/N Error
CAL E
EC,E20
-CAL E
EC,E21
EN
Description and possible countermeasure
Stability error
Weighing value is unstable and "zero display", "sensitivity
adjustment", "calibration test", etc. cannot be executed. Check
around the pan. Refer to the separate instruction manual listed in "1.1
Detailed manuals". Improve the environment of the installation
location (vibration, drafts, static electricity, etc.).
Out of the setting range
The value to be set exceeds the setting range. Set again within the
setting range.
Malfunction of the internal memory element of the balance
If this error continues to be displayed, repair is necessary. Please
contact your local dealer for repair.
Internal weight error
Applying the internal weight does not yield a change in the mass
value as specified. Confirm that there is nothing on the pan and
perform the operation from the beginning.
Internal weight error
The internal weight application mechanism does not function
properly. Perform the operation from the beginning.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary.
Please contact your local dealer for repair.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary.
Please contact your local dealer for repair.
Weighing unit error
Data from the weighing unit has not been received. Check the
connection between the weighing unit and the display unit, and turn
the power off and then on again.
Breeze break unit error
Communication with the breeze break is not possible. Unplug the
power supply and reconnect the cable between the breeze break unit
and the weighing unit.
Serial number error
The serial numbers of the weighing unit and the display unit do not
match. Reconnect with the correct combination.
Calibration weight error (Positive value)
The calibration weight is too heavy. Check around the pan. Check the
calibration mass value.
Calibration weight error (Negative value)
The calibration weight is too light. Check around the pan. Check the
calibration mass value
15

Publicité

loading

Ce manuel est également adapté pour:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte