CW-11 Indicator
4. OPERATION (Cont.)
Gross
Brutto
*
Center
Pcs
of Zero
Net
Tare
Net
Net
Gross
Brutto
4.5 Gross/Net/Tare Recall Operation
When a container has been placed on the platform
and tared, it's weight is stored in memory. Adding
material to the container is shown as NET weight. The
gross weight is a combination of the tared weight and
the material. The G/N/T/ MENU button allows switch-
ing between GROSS, NET and TARE weights.
Repeately press (short presses) the G/N/T/ MENU
button to cycle through Gross, Tare and Net
readings.The sample illustrations indicate a tare
weight of 2kg simulating a container, a net weight
of 8kg which would be the material in a container
and a gross weight of 10kg which is the total weight
of the container and material. After 3 seconds,
display returns to Net weight.
4.6 Unit Switch Operation
To switch measuring units, proceed as follows:
Press and hold PRINT/ UNITS button until display
changes to selected measuring unit. Depending on
which units are enabled in the menu, you have a
choice of g, lb, kg or oz. The display sample
indicates 8kg load changed to lbs shown as a net
weight because a tared weight of 2kg was used
and stored in memory.
4.7 Parts Counting Operation
Parts counting is enabled only when turned ON in the
Readout menu. Refer to paragraph 2.8.4. In the
parts counting mode, the Indicator displays the
quanitity of parts placed on the platform. The Indica-
tor determines the quanitity based on the average
piece weight of a sample. All parts must be reason-
ably uniform in weight for accurate measurements.
EN-29