Write distance unit
Command
Response
aXWDU<value1>! a<value1><value2> <CR><LF>
Example:
Command
Response
0XWDU+0!
0+0+000<CR><LF>
0XWDU+4!
0+0+136<CR><LF>
Read temperature unit
Command
Response
aXRTU!
a<value1><CR><LF>
Example:
Command
Response
aXRTU!
0+0<CR><LF>
Write temperature unit
Command
Response
aXWTU<value1>!
a<value1><value2> <CR><LF>
Example:
Command
Response
0XWTU+0!
0+0+000<CR><LF>
0XWDU+4!
0+0+136<CR><LF>
VEGAPULS C 22 • SDI-12
Description
a: sensor address
<value1>: distance unit +0 = unit in [m], +1 = unit in [ft],
+2 = unit in [mm], +3 = unit in [in]
<value2>: VVO-Status +eee
<CR><LF>
Description
Valid data
No valid data
Current value is returned with a status 136
Description
a: sensor address
<value1>: temperature unit +0 = unit in [°C], +1 = unit in
[F], +2 = unit in [K]
<CR><LF>
Description
Description
a: sensor address
<value1>: temperature unit +0 = unit in [°C], +1 = unit in
[F], +2 = unit in [K]
<value2>: VVO-Status +eee
<CR><LF>
Description
Valid data
No valid data
Current value is returned with a status 136