Extended Commands - Vega PULS C 22 Mise En Service

Masquer les pouces Voir aussi pour PULS C 22:
Table des Matières

Publicité

Command
Response
aD0!
a<value1><value2><value3><va-
lue4><value5><CR><LF>
Measurement data and Format:
Stage value
15)
– +ss.sss (m)
– +ss.sss (ft)
– +ssss.s (mm)
– +sss.ss (in)
Distance
– +dd.ddd (m)
– +dd.ddd (ft)
– +dddd.d (mm)
– +ddd.dd (in)
Electronics temperature
– ttt.t (°C, °F, K)
Measurement reliability
– rrr.r (dB)
Device status
– eee (Errror code)
Example
Address = 0, Stage = 29,272m, Distance = 0,728m, Temperature = 25,4°C, reliability = 14,0 dB,
Device Status = Good
Response : 0+29.272+0.728+25.4+14.0+0<CR><LF>
Address = 4, Stage = 14,887m, Distance = 0,113m, Temperature = 22,7°C, reliability = 14,0dB,
Device Status = M507 (Error in the device setting)
Response : 4+14.887+0.113+22.7+14.0+507<CR><LF>

13.4 Extended Commands

Command
aXRPOM!
aXWPOM<value1>!
aXRDU!
aXWDU<value1>!
aXRTU!
aXWTU<value1>!
The Stage Value outputs the level/gauge above the Stage Reference
15)
VEGAPULS C 22 • SDI-12
Response
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><CR><LF>
a<value1><value2><CR><LF>
a<value1><CR><LF>
a<value1><CR><LF>
Description
Send data (after aM!)
a: sensor address
<value1>: stage value:pss.sss
<value2>: distance value pdd.ddd
<value3>: electronics temperature pttt.t
<value4>: Measurement reliability prrr.r
<value5>: Device status eee
<CR><LF>
Description
Read Power Operation Mode
Write Power Operation Mode
Read distance unit
Write distance unit
Read temperature unit
Write temperature unit
13 Annexe
53

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sdi-12

Table des Matières