ROUTINE CLEANING AND MAINTENANCE
 
In  order  to  maintain  the  performance  and  efficiency  of  the  lamp  germicide,  it  is 
important to strictly comply with the maintenance program attached to the device. 
 
Do not use water jets or sprays for cleaning,  
Use a brush to remove dust build‐up on the air inlet and outlet vent at least every month.  
Use a soft moist cloth to remove dust build‐up from the structure. 
 
BEFORE EVERY INTERNAL MAINTENANCE OPERATION UNPLUG THE DEVICE! 
 
 
Every 3 months (approximately 2,000 hours) replace the filter inside the air 
inlet vent (Spare filter: Art. SSAA015). 
 
 
Follow these instructions step‐by‐step to replace the filter:  
1. Take the supplied L‐shaped key ( 3MECL0001) specific for 
TORX safety screws (Fg.6) and unscrew the front part; 
2. Remove the front of the vent where the filter is housed.     
(Fg.7) 
3. Replace the filter (SSAA015). 
4. Put the front of the vent back on in the order shown in Fg.7.  
 
 
 
FILTRO: SSAA015 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
16 
Fg.6 
Fg.7 