10 Menu
10.1 Navigation in the menu
Call up menu
Select function
Change settings
Exit menu/
Return to
weighing mode
28
Access to service menu „tCH"is locked in verified balances.
To disable the access lock, destroy the seal and actuate the
adjustment switch. For position of adjustment switch, see
chap. 13.
Attention:
After destruction of the seal the weighing system must be re-
verified by an authorised agency and a new verification
wire/seal mark fitted before it can be reused for applications
subject to verification.
Turn on the scale during the self-test press
function [F1 oFF] is displayed.
With help of
, the individual functions can be selected
one after the other.
Confirm selected function by
be displayed.
Select the desired setting with
confirm or
to reject, the balance returns to the menu.
Press
, the balance will return to weighing mode.
, the first
. The current setting will
and press
to
MBC-M-BA-e-1615