Environmental
Operating Condition
Transport and Storage
Condition
Dimension and Weighting
Weight (without batteries)
Size
Compliance
Item
Equipment classification
Type of protection
Degree of protection
Front panel and case
labeling
Temperature
Home healthcare
environment
Safety classification of ME EQUIPMENT
Protection against electric shock
Applied part
Protection against harmful ingress of
water or particulate matter
Mode of operation
Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT
ENGLISH
Operating Temperature: 15℃~40℃
(59°F ~104°F),
Relative Humidity ≤85%,
atmospheric pressure:70 Kpa -106 Kpa
Storage Temperature:-20°C -55°C /
(-4°F - 131°F),
Relative Humidity ≤93%,
Atmospheric pressure:70 Kpa -106 Kpa
80g
L:122mm X W:96mm X H:42mm
Compliant with
Safety Standards: EN 60601-1: 2006+A1:2013,
EN 60601-1-2: 2015
Internally powered equipment (on battery power)
Non Applied part
EN ISO15223-1:2016
EN ISO80601-2-56:2017
EN 60601-1-11:2015
Internally powered ME equipment
Non Applied part
IP22
Continuous operation