Important
Do not boil or autoclave the masks and air hose.
• Once all components are fully dry, reassemble the
atomiser and place the components in a dry, sealed
container.
Note
Ensure that all parts are thoroughly dried during, and
in particular after cleaning, otherwise there is an in-
creased risk of bacterial growth.
Drying
• Place the individual components on a dry, clean and
absorbent surface and allow to dry completely (for
at least 4 hours).
Material resistance
• As with all plastic components, frequent use and
cleaning of the atomiser and the accessories can
lead to a certain amount of wear. Over time, this can
alter the aerosol properties and may eventually affect
the efficiency of therapy. We therefore recommend
that you replace the atomiser once a year.
• When choosing a cleaning or disinfecting agent, note
the following: Only use mild cleaning or disinfect-
ant agents in the quantities recommended by the
manufacturer.
Storage
• Do not store the nebuliser in a damp atmosphere
(e.g. in the bathroom) or transport it together with
damp objects.
• Store and transport the nebuliser away from direct
sunlight.
• The accessories can be safely stored in the acces-
sories compartment [4]. Store the device in a dry
place, if possible in its original packaging.

9. Disposal

In the interest of environmental protection, do not dis-
pose of the device in household waste.
Please dispose of the device in accordance with
EC Directive – WEEE (Waste Electrical and Elec-
tronic Equipment).
If you have any queries, please contact the appropriate
local authorities.

10. Troubleshooting

Problem/
Possible Cause/Remedy
Question
The atomiser
1. Too much or too little medicine
produces
little or no
aerosol.
2. Check that the nozzles are not
3. The atomiser is not held in an
4. The medicine solution is
Output is too
Hose kinked, filter blocked, too
low.
much inhalation solution.
Which
Consult your doctor.
medicines are
Essentially all medicines that
suitable for
are suitable and approved for
inhalation?
device-based inhalation can be
inhaled.
Some inhalation
This is normal and occurs for
solution
technical reasons. Stop inhala-
remains in the
tion if you hear a notable differ-
atomiser.
ence in the noise of the atomiser.
What special
1. In babies, the mask should
steps should be
taken for babies
and children?
2. In children, the mask should
Note: Children should only use
the device with help and under
supervision of an adult. Never
leave a child alone with the
nebuliser.
Inhalation with
This is for technical reasons.
the mask takes
Less medicine is inhaled per
longer.
breath through the mask holes
than using the mouthpiece. The
aerosol is mixed with ambient air
through the holes in the mask.
18
in the atomiser.
Minimum: 2 ml,
Maximum: 8 ml.
blocked If necessary, clean the
nozzle (e.g. by rinsing). Use the
atomiser again.
CAUTION: Only poke imple-
ments through the fine holes
from the underside of the nozzle.
upright position.
unsuitable for nebulising (e.g.
too thick).
The medicine solution should
be specified by the doctor.
cover the mouth and nose to
guarantee effective inhalation.
also cover the mouth and
nose. Nebulisation next to a
sleeping person is not suitable
because insufficient medicine
reaches the lungs.