VIII. Maintenance and Cleaning
1. To guarantee trouble-free use for many years it is necessary to clean the dispenser
regularly. If there is any sign of a potential malfunction (e.g. piston is difficult to move)
never use force. Immediate cleaning should be carried out.
2. Do not leave the dispenser empty without cleaning to avoid clogging the valve balls!
If this does happen soak the dispenser for some hours in cold water or clean it using
ultrasound. Once the piston can be moved, dismantle and remove any residual impurities
by cleaning with a normal solution.
Disassembly of the Dispenser:
a) Remove the intake tube.
b) Unscrew the discharge tube (caution- do not remove the thread nipple from the
valve block)
c) Pull the finger protection cap up to the top, loosen the volume setting and remove
it by pulling it down
d) Pull the finger protection cap and dispensing piston out of the dispensing piston
3. When changing the dispensing liquid the dispenser should be cleaned.
Please note: When mounting the
screwed into the glass cylinder carefully otherwise the piston and cylinder can get
jammed.
IX. Repair Service
1. All OPTIFIX
dispensers are made from highest quality materials and are almost
®
maintenance-free and reliable. If a repair should be necessary, it must be done by the
manufacturer or an authorized service station. This does not refer to parts available
as spare parts (finger guard with piston, volume setting system, discharge tube set,
tubes, adaptors).
2. Before returning the dispenser please do not forget to register it with the
manufacturer in advance. Make sure that it is perfectly cleaned to avoid any
contamination. Please also enclose a filled in safety declaration.
3. Please use the dispenser following to the operating manual to be entitled to the war-
ranty. Repairs must be carried out by the manufacturer or authorized service stations.
4. Please note that the
changeable.
X. Disposal
When disposing the dispensers please pay attention to the national disposal regulations.
XI. Technical Data (EN ISO 8655/5)
1. Volume Range - Subdivisions - Accuracy
Volume range
in ml
0,5 - 2,0
1,0 - 5,0
2,0 - 10,0
5,0 - 30,0
10,0 - 50,0
20,0 - 100,0
40,0 - 200,0
60,0 - 300,0 *
100,0 - 500,0 *
*not acc. to ISO 8655/5
SOLVENT
HF
dispensing cylinders and dispensing pistons are not inter-
Graduation
in ml
0,1
0,1
0,2
0,5
1,0
2,0
5,0
5,0
5,0
Dispenser the glass piston has to be
Accuracy
Precision
A in %
< + 0,6
< + 0,6
< + 0,6
< + 0,6
< + 0,6
< + 0,6
< + 0,6
< + 0,6
< + 0,6
CV in %
< + 0,2
< + 0,2
< + 0,2
< + 0,2
< + 0,2
< + 0,2
< + 0,2
< + 0,2
< + 0,2
17–56