5.4 Battery operation
The bed is equipped with a battery, which means it can be adjusted independently from the
supply network.
The number of adjustments depends on the charge status of the battery. Complete dischar-
ging of the battery significantly reduces its service life. To prevent the batteries from excessive
discharge, the bed should remain connected to the power supply. The battery charges auto-
matically via an integrated charging board
– The battery is maintenance-free
– It is integrated in the power box
– It is a lead gel battery which can be operated irrespective of its location.
– Nominal data: 24V, 1.2Ah
– Depending on the usage, the service life of the battery is approx. 5 years
– Disposal is only permitted at special collection points.
Attention!
Harmful chemicals can leak from opened batteries and cause injuries. Never open
the battery box.
Exchanging the batteries may only be performed by trained personnel. The service life of the
battery strongly depends on its use. Complete discharging of the battery significantly reduces
its service life.
5.5 Moving and braking (sentida 7-i | cozy)
The bed can be moved and therefore the easy positioning within the room ensures easy clea-
ning of the nursing bed and the space underneath. However, the nursing bed is not suitable
for the transport of residents and patients. The nursing bed may only be moved in its lowest
position and without a patient. Moving the nursing bed to other rooms with a resident or pati-
ent still in it dow not comply with the intended use.
There is a covered axle with lockable castors beneath the head and foot end. On the inner side
of this undercarriage cover, there is a braking lever located at the left and right side of the bed.
With this lever, you can release or lock the castors for each axle.
Caution!
The bed can only be moved if both brakes are released. If braked castors are moved
across the floor, the flooring could be damaged.
73