5.3.5.4 Load default values
Menu path:
SERVICE / DEFAULTS / LOAD DEFAULT
Description:
This command loads all default values of the instrument.
After loading the default values the parameters can freely be changed, the effect of the changes does not affect the measurement until
the user exits programming mode and returns to measurement mode. Before loading the defaults the software asks for a confirmation
warning the user that all user parameters will be lost!
5.3.5.5 Service distance offset
Parameter:
P05
Menu path:
SERVICE / SERVICE DIST OFFSET
Description:
There is a possibility to display auxiliary service information on the bottom line of the screen.
This information is useful when verifying measurement is performed with a hand-instrument
and zero point of this device is not the same as the highest position of the float. In this case
a distance should be entered into this Parameter which is the distance between the highest
position of the float (which is the zero point of the measurement range) and the zero point of
the verifying instrument.
This Parameter has no effect on level measurement or on volume and mass calculation, it
only appears on the screen. If the value of this Parameter is not zero, „SDIST=x.xxx" format
display appears on the bottom line of the measurement screen.
5.3.5.6 Restart
Menu path:
SERVICE / RESTART
Description:
Restarts the instrument (Cold boot) (Reloads parameters from the non-volatile memory)
FM16CA0037X ♦ FM16US0068X ♦ IEC Ex BKI 12.0002 ♦ BKI16ATEX0012X ♦ mba5052i0600p_06 ♦ 33 / 36
Default value:
0 mm