7.1 TO USE THE CLAMP-ON METER
Set the selector (3) to any position other than OFF.
7.2 TO STOP THE CLAMP-ON METER
Set the selector (3) to OFF.
Automatic shutdown
The multimeter shuts down automatically 10 minutes after the last
measurement; the buzzer sounds one minute before shutdown.
Pressing any button except SELECT restarts the multimeter.
Cancelling automatic shutdown (CM625 only)
1.
Set the selector (3) to OFF.
2.
Hold down the HOLD button and set the selector (3) to any
position other than OFF.
The
deactivated.
Restarting automatic shutdown (CM625 only)
1.
Set the selector (3) to OFF.
2.
Set the selector (3) to any position other than OFF.
The
reactivated.
7.3 TO MEASURE VOLTAGE
Please read the safety recommendations before use.
Automatic measurement ranges:
-
CM600 and CM610: 200 mVDC to 600 VDC (5 ranges) and
2 VAC to 600 VAC (4 ranges)
-
CM625: 4 VDC to 600 VDC (4 ranges) and 4 VAC to 600 VAC
(4 ranges)
1.
Set the selector (3) to the Voltage symbol.
2.
Insert the black connector into the COM terminal (1), the
red connector into the + terminal (10) and take the voltage
reading once it has stabilised.
When the AUTO icon (23) is displayed, the multimeter
determines the optimum measurement range.
If "-----" is displayed, the capacity has been exceeded.
- CM600 and CM610: to display the maximum value, see §. 7.9.
- CM625: to store the maximum reading, see §. 7.10.
3.
Set the selector to OFF.
7.4 TO MEASURE DC CURRENT (CM625)
Please read the safety recommendations before use.
Automatic measurement ranges:
-
CM625: 40 ADC or 400 ADC (2 ranges)
1.
Set the commutator (3) to Zero
2.
Press the SELECT button (9) to set zero.
Setting zero can only be activated on the 40 A range.
3.
Press the trigger (4) to open the clamp (6) and position the
conductor in the centre of the clamp (Fig. 4).
NB: never insert two + and – conductors from the same circuit
in the clamp; the reading will be zero.
4.
Take the current reading once it has stabilised.
If "-----" is displayed, the capacity has been exceeded.
To store the maximum reading, see §. 7.10.
5.
Open the clamp, release the conductor and set the selector
to OFF.
7.5 TO MEASURE THE AC CURRENT
Automatic measurement ranges:
-
CM600 and CM610: 200 AAC or 600 AAC (2 ranges)
-
CM625: 40 AAC or 400 AAC (2 ranges)
1.
Set the selector (3) to
The icons (15 and 21)
2.
Press the trigger (4) to open
conductor in the centre of the clamp (29).
NB: never insert two + and – conductors from the same circuit
in the clamp; the reading will be zero.
3.
Take the current reading once it has stabilised.
If "-----" is displayed, the capacity has been exceeded.
- CM600 & CM610: to display the maximum value, see §. 7.9.
- CM600 & CM610: to store the maximum reading, see §. 7.10.
4.
Open the clamp, release the conductor and set the selector
to OFF.
7. USE
icon disappears and automatic shutdown is
icon appears and automatic shutdown is
.
app
ear
.
12
.
the clamp (6) and position a