●Loosen the pump securing screw which on the bottom of control station.
(See below pictures)
●Insert the power plug into the socket
●Turn on the power switch and the lamp will be lit
●Don't pull out the power plug instantly after turning off the power switch,
because the fan keeps operating to protect the heating element. The
plug can be removed once the fan has stopped completely.
III. Calibrating the iron and hot air temperature
The soldering iron and hot air gun should be recalibrated after changing the iron/gun,
or replacing the heating element or tip/nozzle.
1.
Connect the cord assembly plug to the socket on the station.
2.
Set the temperature control the knob to 400°C (750°F).
3.
Turn the power switch to 'ON' wait until the temperature stabilizes, then remove
the CAL pot plug.
When the temperature stabilizes, use a straight-edge(-) screwdriver or small plus(+)
screwdriver to adjust the screw (marked CAL on the station) until the tip thermometer
indicates a temperature of 400°C(750°F). Turn the screw clockwise to increase the
temperature and counterclockwise to reduce the temperature. Replace the CAL pot
plug.
IV. Operation instructions (SMD rework)
●Remove SMD components (such as QFP, SOP, PLCC and so on)
EN-5