Paramètres Profibus-Pa - Endress+Hauser mycom CLM 152 Instructions De Montage Et Mise En Service

Masquer les pouces Voir aussi pour mycom CLM 152:
Table des Matières

Publicité

Interface Profibus
68
10.7
Paramètres Profibus-PA
10.7.1 Paramètres PROFIBUS-PA inductif
Paramètre
Composite List Directory
DEVICE_ID
Actual Error
Device Bus Address
Device and Software
Number
Main measured value
Temp. measured value
Ext. binary input
Operating mode
Measuring cell
Input damping
Set ATC/MTC
MTC temperature
Cell constant
Adaptation factor
Temp. offset
Select substance MR 1
Select substance MR 2
Select substance MR 3
Select substance MR 4
Instrument number
Software version
Rest counter
SAP code
Measuring point
description
PNO ID number
Matrice
Indice
Type
VH
(Slot = 1)
données
-
1
Octet String
V99H0
25
Octet String
90
42
Unsigned 16
94
44
Integer 8
93
48
Unsigned 16
00
108
Float
01
109
Float
05
110
Unsigned 8
06
111
Unsigned 8
07
112
Unsigned 8
10
113
Unsigned 8
11
114
Unsigned 8
13
115
Float
22
116
Float
24
117
Float
28
118
Float
60
119
Unsigned 8
61
120
Unsigned 8
62
121
Unsigned 8
63
122
Unsigned 8
91
123
Unsigned 32
92
124
Unsigned 16
95
125
Unsigned 8
96
126
Octet String
A0
127
Octet String
A1
128
Unsigned 16
mycom CLM 152
Lire
Ecrire
Longueur
donnée
yes
24
yes
16
yes
2
yes
1
yes
2
yes
4
yes
4
yes
1
yes
1
yes
1
yes
yes
1
yes
yes
1
yes
yes
4
yes
4
yes
yes
4
yes
4
yes
yes
1
yes
yes
1
yes
yes
1
yes
yes
1
yes
4
yes
2
yes
1
yes
18
yes
yes
32
yes
2
Endress+Hauser

Publicité

Table des Matières
loading

Table des Matières