REAR PANEL
3 )
RJ45 opto mos relay output connector corresponding to alarm level exceeding.
GPO1
GPO2
Internal OPTO MOS wiring
CMD 2
CMD 1
4 )
RJ45 connector to ethernet (LAN)
5 )
B type USB connector for firmware upgrade only.
Caution: Do not connect any computer to this jack, excess risk of
damage to your network adapter.
MEASURING RANGE
With A weighted, Leq level range is from 60 dB(A) to 122 dB(A).
However, it is possible, as an option, to measure from 40 dB(A) to 102dB(A).
With C weighted,Leq level range is from 60 dB(A) to 122 dB(A), and peak level range is from
70 dB(C) to 132 dB(C).
However, it is possible, as an option, to measure from 50 dB(C) to 112 dB(C).
ERROR LISTING
Error listing which can be displayed by control display (B).
E2 : Sensor link defect.
E3 : Low internal battery voltage. Replace quickly.
ENGLISH
1
+
SENSOR
-
V+
OPTO MOS 2
OPTO MOS 1
8
1
2
3
V+
4
5
6
7
8
Leds GPO1 and GPO2 display GPO1 and GPO2
state, and allow to recover the equipment IP address.
(see page 37)
Note :
Each OPTO can be considered as a relay and
offers all the advantages (alternative signals, no
wiring). Current limit is fixed at 200mA.
34
Next
AMIX - AFFseries-2