C
L
YCLE
OCK
To ensure repeatability, the centrifuge can be locked either on one cycle (Single Cycle Lock) or restricted to the saved
cycles (Presets Lock). The Single Cycle Lock also prevents making changes to the selected cycle parameters.
The Preset Lock allows selection of any saved cycle and prevents changing the parameters of saved cycles.
S
C
L
INGLE
YCLE
OCK
Lock on Single Cycle
Cancel Single Cycle
P
L
RESET
OCK
Enter Settings
Lock Saved Cycles
Cancel Preset Lock
Select desired cycle. With lid open, press and hold the UNLOCK
button for 5 seconds. Two beeps and a flash of the lid light will
confirm that cycle selection is locked.
To re-enable cycle selection, with lid open, press and hold the
UNLOCK button for 5 seconds. Two beeps and a flash of the lid
Lock
light will confirm that cycle selection is now unlocked.
Press the GEAR button.
Mode
With lid open, press and hold the UNLOCK button for 5 seconds.
Two beeps and a flash of the lid light will confirm that the Preset
Lock is active. The cycle screen will now only show the numbers
for programmed cycles.
With lid open and while not in Settings mode, press and hold the
UNLOCK button for 5 seconds until two beeps are heard. Saved
cycles can again be edited or programmed.