Settings; Standard Settings; Review Cycle Time And Speed Settings; Changing Cycle Time And Speed Settings - Drucker Diagnostics HORIZON 24-AV Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
S
ETTINGS
S
S
TANDARD
ETTINGS
RPM
Time
R
C
T
EVIEW
YCLE
IME AND
Factory programmed cycles are shown on the rear of the centrifuge, on the Factory Set Cycles label. To review
current settings, follow this procedure:
o The lid must be open to review the selected cycle time and speed.
o Press and hold the START button until you hear a beep.
o Release the START button. The centrifuge will beep and the LED light will flash once for each minute of run time
in the current cycle. 10 beeps / flashes equal 10 minutes of run time. Run time starts when the rotor reaches 90%
of desired speed and stops when the rotor starts decelerating.
o Pressing the START button again will cause the unit to beep and the LED light to flash once for each 100 rpm in
the current cycle. 10 beeps / flashes equal 10 x 100 or 1,000 Revolutions per Minute (RPM).
o The centrifuge will automatically revert to normal mode at the end.
C
C
T
HANGING
YCLE
IME AND
o The lid must be open to change the selected cycle time and speed.
o Select the cycle you wish to change.
o Press and hold the START and UNLOCK buttons together until the LED light flashes.
o Press the START button for each minute of run time.
o Move to speed setting mode by pressing the UNLOCK button.
o Press the START button once for each 100 rpm.
o Press the UNLOCK button to exit setting mode.
Setting A
1,500
5 min
S
S
PEED
ETTINGS
S
S
PEED
ETTINGS
Drucker Diagnostics | +1-814-692-7661 |
Setting B
2,500
4 min
CustomerService@DruckerDiagnostics.com
Setting C
1,500
4 min
| 5

Publicité

Table des Matières
loading

Table des Matières