AC Compressor Atomiser – Guidance and manufacturer's declaration
Table 2 – Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
( 5 0/60Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
Compliance level
Test level
±8 kV contact
±8 kV contact
±15 kV air
±15 kV air
Power supply
Power supply lines
lines ±2 kV
±2 kV
input/output
lines ±1 kV
line(s) to line(s)
line(s) to line(s)
±1 kV.
±1 kV.
line(s) to earth
100 kHz repetition
±2 kV.
frequency
100 kHz repeti-
tion frequency
5% U
5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5 cycle
for 0.5 cycle
40% U
40% U
T
( 6 0% dip in U
)
( 6 0% dip in U
T
for 5 cycle
for 5 cycle
70% U
70% U
T
( 3 0% dip in U
)
( 3 0% dip in U
T
for 25 cycle
for 25 cycle
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5s
for 5s
30 A/m
30 A/m
50Hz/60Hz
50Hz/60Hz
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic ti-
le. If floors are covered with synthetic material,
the relative humidity should be at least 30%
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
T
)
commercial or hospital environment.
T
T
)
T
T
)
T
T
)
T
Power frequency magnetic fields should be
at levels characteristic of a typical location in a
typical commercial or hospital environment.
27