Technical Data - Zeiss WSB ZPiezo CAN Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

WSB ZPiezo CAN
2.3

Technical Data

Dimensions
Weight
Ambient conditions
Storage (in its packaging)
Permissible ambient temperature
Permissible relative atmospheric humidity
Transport (in its packaging)
Permissible ambient temperature
Operation
Permissible ambient temperature
Permissible relative atmospheric humidity
Air pressure
Degree of contamination
Operating area
Height of the operating area
Technical Operating Data
max. Z travel range with WSB 500 stage attatchment
Resolution
Repeatability
Resonant frequency @250g
Settling time 10-90%
Recommended warm-up period
Protection class
Protection type
Electrical safety
Degree of contamination
Overvoltage category
Mains voltage
Mains frequency
Power consumption: Table-top power supply unit with connected controller
Table-top power supply unit output
Table-top power supply unit input
07/2015
DEVICE DESCRIPTION
Technical Data
(Width x Depth x Height) 252 mm x 221 mm x 91 mm
Due to the wide operating voltage range of the device power supply unit,
432339-9010-000
maximal 75 % at +35 °C (non-condensing)
to DIN EN 61010-1 (IEC 61010-1)
in compliance with CSA and UL regulations
no conversion of the mains voltage is required!
100-240 V, 50-60 Hz, max. 0.4 A
Carl ZEISS
2.5 kg
+10 to +40 °C
-40 to +70 °C
+10 to +40 °C
maximum 75 %
800 hPa to 1060 hPa
2
enclosed rooms
max. 2000 m
500 µm
1 nm
±5 nm
170 Hz ± 20 %
9 to 16 ms
15 minutes
II
IP 20
2
II
100 to 240 V ±10 %
50 Hz – 60 Hz
max. 50 VA
24 V DC, max. 1 A
9