Télécharger Imprimer la page

ResMed AirMini Guide De L'utilisateur page 30

Masquer les pouces Voir aussi pour AirMini:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 74
Declaration of Conformity (DoC to the Radio Equipment Directive)
ResMed declares that the AirMini device (models 381xx) is in compliance with the
essential requirements and other relevant provisions of Directive 2014/53/EU (RED). A
copy of the Declaration of Conformity (DoC) can be found on
Resmed.com/productsupport
This radio equipment operates with the following frequency bands and maximum radio-
frequency power: Bluetooth Class 2. 2402 to 2480 MHz, 4dBM.
Aircraft use
ResMed confirms that machine meets the Federal Aviation Administration (FAA)
requirements (RTCA/DO-160, section 21, category M) for all phases of air travel.
Wireless technology
Technology used:
Connection types:
Frequency:
Max RF power output:
Operating range:
It is recommended that the machine is at a minimum distance of 1.1cm from the body
during operation. Not applicable to masks, air tubing or accessories.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
Operating pressure range
AutoSet, AutoSet For Her, CPAP:
Maximum single fault steady pressure
Machine will shut down in the presence of a single fault if the steady state pressure
exceeds 30 cm H
2
than 1 second.
28
O (30 hPa) for more than 6 seconds or 40 cm H
Bluetooth
SPP, iAP2, GATT
2402 to 2480 MHz
+4 dBm
10 m (Class 2)
4 to 20 cm H
O (4 to 20 hPa)
2
O (40 hPa) for more
2

Publicité

loading