HBM PACEline CMC Manuel D'emploi page 17

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CMC
RANGE1 / RANGE 2
If a voltage of 0 V is present at Pin 2, measuring range 1 (100% F
active at the charge amplifier. The charge amplifier gives you the option to
zoom into a second measuring range (20 % of the nominal (rated) force /
output span). A voltage of 24 V DC must be applied at PIN 2 for this.
10 V
2 V
0 V
0%
RANGE 1 / RESET
Fig. 6.3
Zooming into the second measuring range
Changing the measuring range to 20% can also be used to operate the force
transducer with high safety reserves for measurement tasks where overload is
critical. The charge amplifier overloads above 20%.
Supply voltage 0 V (GND)
RANGE 1 / RANGE 2
RESET / MEASURE
TEDS
Output signal -10 - +10 V
Signal ground
Do not assign !
Supply voltage 18 - 30 V
M12 x 1, 8‐pol
1)
Operation with SELV circuit (separated extra low voltage)
Fig. 6.4
PIN assignment
A2319-4.0 en/de/fr
100%
1
2
3
4
5
6
7
8
1)
DC
The cable shield must not be connected to the
housing ground of the downstream evaluating
electronics. PIN 7 is for HBM internal use only and
must not be assigned.
10 V
0 V
0%
20%
RANGE 2 / RESET
NOTE
17
) is
nom
100%
white
brown
green
Connection
yellow
cable wire
gray
assignments
KAB 168
pink
blue
red
HBM

Publicité

Table des Matières
loading

Table des Matières