HBM PACEline CMC Manuel D'emploi page 16

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

16
Setting RESET will set the amplifier output signal to zero. This can be done
with any applied force. The advantage of the Reset function is that you can
compensate initial loads and drift factors. A measurement can also be started
at a high force level.
10 V
2 V
0 V
0%
-8 V
Fig. 6.2
Reset function
ATTENTION
After RESET, although the charge amplifier output is set to
zero, this does not mean that the machine generating the
force is force‐free.
Make sure that the force transducer is not overloaded, even
though the output signal is still in the -10 ‐ +10 V range.
Once all the force is removed, a negative voltage signal is
present at the value of the voltage output for a RESET,
without having to reset again.
HBM
RESET
É É É É É É É É É É
É É É É É É É É É É
É É É É É É É É É É
OVERLOAD RANGE
É É É É É É É É É É
É É É É É É É É É É
É É É É É É É É É É
80%
100% F
nom
Removal of
force
CMC
A2319-4.0 en/de/fr