SEnxxx – Set Efficiency
This command sets the alpha or beta efficiency. The efficiency must be
multiplied by 10 before sending.
1 = Set alpha efficiency.
2 = Set beta efficiency.
SHnxxxxxxxxxxxxxxx – Set Header
This command sets a user-defined header. There are six headers accessed 1-
6. Each header is 15 characters long.
SLyyyyy.y – Set Loss of Count Timer
This command sets the loss of count timer in seconds. The maximum is
99999.9 seconds. A time of 0 will disable this function.
SNnyyyyyy – Set MDA
This command sets the alpha and beta MDA.
1 = Set alpha MDA.
2 = Set beta MDA.

SPxxxxx.x

This command sets the user-defined count time when the COUNT TIME
SWITCH is in the PC position. The maximum is 99999.9 seconds.
SQnn – Set QC parameters
This command sets the specified QC parameter.
01 = Set QC enabled status, 0 = OFF, 1 = ON.
04 = Set standard alpha efficiency %., format ##
05 = Set standard beta efficiency %, format ##
06 = Set alpha source size (cpm), format ######
07 = Set beta source size (cpm), format ######
08 = Set alpha efficiency limit %, format ##
09 = Set beta efficiency limit %, format ##
10 = Set alpha background upper (cpm), format ######
11 = Set alpha background lower limit (cpm), format ######
12 = Set beta background upper (cpm), format ######
13 = Set beta background lower limit (cpm), format ######
17 = Set last QC time, format HHMM
18 = Set QC Mode, 0 = NOR, 1 = BKG
19 = Set QC Update Mode, 0 = OFF, 1 = ON, 2 = EFF, 3 = SUB
Manuel Technique
Page 37