Commandes (Cette Partie N'est Pas Traduite); Bn - Background Subtract Status; Cn - Count; Dn - Display Mode - Ludlum Measurements 3030P Manuel Technique

Table des Matières

Publicité

Modèle 3030P
Ludlum Measurements, Inc.
COMMANDES
Bn – Background Subtract Status
This command sets background subtract ON/OFF or return background
subtract status, where "n" is a value of 0, 1, or 2.
0 = Sets background subtract OFF.
1 = Sets background subtract ON.
2 = Returns status.

Cn - Count

This command performs various count functions. The count time is
determined by the positioning of the COUNT TIME SWITCH on the front
panel unless otherwise noted. See command RR for the format of the count
results.
1 = Start a count.
2 = Stop a count in progress.
3 = Clear alarms.
4 = Start a count and automatically return the results.
5 = Start a one-minute count and automatically return the results.
Dn – Display Mode
The display can be set to display count data as raw scaler counters, counts
per minute (CPM), and disintegrations per minute (DPM).
0 = Set display mode to scaler.
1 = Set display mode to CPM.
2 = Set display Mode to DPM.
3 = Return display mode (SCA, CPM, or DPM).
F – Firmware
This command returns a string containing the firmware version number.
Ln – Logging Mode
This command sets how samples are logged to internal memory.
0 = Turn logging OFF.
1 = Turn Logging ON.
2 = Log only QC checks.
3 = Return logging mode (OFF, ON, or QC).
Manuel Technique
Page 32
Section 7
May 2013

Publicité

Table des Matières
loading

Table des Matières