Rr - Read Count; Rs - Read Samples; Rt - Read Current Count Time; Ru - Read Comment - Ludlum Measurements 3030P Manuel Technique

Table des Matières

Publicité

Modèle 3030P
Ludlum Measurements, Inc.
17 = Reads last QC time. Returns "HH:MM."
18 = Reads QC Mode. Returns "NOR" or "BKG."
19 = Reads QC Update Mode. Returns "OFF," "ON," EFF," or
"SUB."
20 = Reads alpha source count time (secs). Returns "nnnnn.n."
21 = Reads beta source count time (secs). Returns "nnnnn.n."
22 = Reads background count time (secs). Returns "nnnnn.n."
23 = Reads QC override status. Returns "OFF" or "ON."
RR – Read Count
This command reads the current alpha and beta counts. Returns "HH:MM
MM/DD/YY aaaaaa bbbbbb x tttt.t."
aaaaaa = alpha count
bbbbbb = beta count
x = S (scaler), C (cpm), and D (dpm)
tttt.t = count time in minutes (If count is in progress, this value
shows count time remaining.)
RS – Read Samples
This command reads all logged samples from memory. While reading
samples, the audio is stopped and the displays show a scrolling series of
dashes along with the sample number. A "$" marks the end of the samples.
Returns: "ssss MM/DD/YY HH:MM:SS aaaaaa bbbbbb ttttt.t x
cccccccccc[CR][LF] - 59 bytes $[CR][LF] - 3 bytes"
ssss = sample number
aaaaaa = alpha scaler
bbbbbb = beta scaler
x = S (Scaler), C (cpm), and D (dpm)
ttttt.t = count time in seconds
cccccccccc = user defined comment
RT – Read Current Count Time
This command reads the current count time in seconds. Returns "nnnnn.n."
RU – Read Comment
This command reads the user-defined, 10-character comment field. Returns
"xxxxxxxxxx."
RV – Read High Voltage
This command reads the high-voltage set point. Returns "nnnn."
Manuel Technique
Page 35
Section 7
May 2013