INTRODUCTION
Metrological maintenance and care information
• Do not perform any metrological maintenance
EN
if the device is in use.
• For reliable operation, we recommend that
the device is subject to metrological control
once every two years pursuant to Sec. 14 of
the Medical Product Operating Directive (MP-
BetriebV). We recommend testing the device's
performance once every two years and after
metrological control and repair, at a minimum
by testing the tolerance requirements of the
pressure display and tightness of the cuff
(testing at least at 50 mmHg and 200 mmHg).
• The metrological control must be performed by
a licensed tester or testing institute.
• We recommend cleaning the device and cuff
only with a soft cloth.
30
Storage information
• If the device is not being used, store it in a dry
location away from heat, dust, fibres, and out of
direct sunlight. Do not place heavy objects on
the storage box.
Information about handling batteries
• Do not mix new and used batteries.
• Do not use different types of batteries at the
same time.
• Do not throw batteries into a fire. Batteries may
explode or leak. Remove the batteries if the
device is not used for a long time.