Note
Before the treatment, ensure that the skin is shaved, clean and dry (free from any residue such as
powder, cream, make-up, antiperspirant or deodorant).
Allow the device to cool for a few minutes if it seems too hot for you. You can continue your treat-
ment as normal after a short cooling period.
WARNING
Do not clean the skin with flammable liquids such as alcohol or acetone shortly before using the
beurer SalonPro System.
5. Flip the power switch on the back of the base unit. The indicator lamp and the energy indicator lamp
are illuminated and the noise of a fan (similar to the noise a hair dryer makes) can be heard.
6. The device is automatically set to the lowest energy level. The device is then ready to emit the first
pulse.
7. The energy selection buttons "+" and "-" can be used to increase or decrease the energy level to set
the energy level determined in the automatic skin tone recognition test and confirmed in the skin sen-
sitivity test and/or to adjust the setting to your requirements. The energy indicator lamps illuminate
green in accordance with the selected energy level.
8. Place the light area of the cartridge on the desired treatment area and make sure it is in full contact
with the skin.
Note
If you press the trigger button and your skin is to dark or too tanned, the indicator lamp on the applicator
illuminates red. In this case, the applicator does not trigger a pulse on this area of skin.
If the applicator does not have sufficient contact with the skin and you press the trigger button, the
indicator lamp on the applicator illuminates orange. In this case, the applicator does not trigger a
pulse on this area of skin either.
9. Press the trigger button. You will see a bright flash, hear a cracking noise and may feel a slight burn-
ing sensation and/or warmth. The beurer SalonPro System features a "flash-glide" function. When
you press and hold the trigger button, a bright flash is automatically triggered every three seconds,
meaning that you do not have to press the trigger button every time.
WARNING
Do not treat the same area of skin more than once per hair removal session!
Treating an area of skin more than once during a single hair removal session can increase the prob-
ability of side effects.
Try to avoid overlapping pulses.
Note
The pulses from the beurer SalonPro System should be applied in rows – beginning at one end of
each row, continuing to the other end. This technique allows better monitoring of the areas of skin
that have already been treated and helps you to avoid treating the same area more than once or
treating overlapping areas of skin. The light area of the beurer SalonPro System is designed in such a
way that it leaves a temporary impression on the treated area of skin. These visible impressions can
be used for precise placement of the next pulse. Additionally, a lengthways marking is located on the
left- and right-hand side of the cartridge that helps you to determine precisely where the light area is
at any time.