OHAUS STARTER 3100C Manuel D'instructions page 17

Appareil de mesure de la conductivité de laboratoire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

STARTER 3100C Bench Conductivity Meter
The details for the second line are:
Auto EP, 85.1μS/cm, 26.8° C, ATC
|
|
|
|
|
|------------------------- Conductivity value
|---------------------------------------- Auto End Point
The details for the third line are:
Ref.25° C , T.Comp.Coeff. 2.00%/° C
|
|
|------------------------------------------- Reference temperature, 20 or 25 ° C
The format for the print-out following a TDS measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 38.0mg/L, 23.2° C , ATC
Ref.25° C , T.Comp.Coeff. 2.00%/° C
TDS-F 0.5
The format for the print-out following a Salinity measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 0.08psu, 25.0° C , MTC
The print-out in case of an error message is:
End Point, Value, Temp., ATC/MTC
Error4
The print-out for a successful calibration is:
Standard: 1413μS/cm
Temperature: 30.0 ° C
Calibration value: 1552μS/cm
Cell Constant: 1.66 cm-1
ATC/MTC: ATC
Printing from memory When scrolling through the memory you can print the entry that is currently
viewed by pressing and holding
Recall 08:
EndPoint, Value, Temp., ATC/MTC
Auto EP, 68.8μS/cm, 26.8° C , ATC
Ref.25° C , T.Comp.Coeff. 2.00%/° C
|
|---- Auto Temp. Compensation
|----------------Temperature value and unit
|
|---------------- Temperature compensation coefficient value
for 2 seconds. The printout format is followings:
EN-13