Using the memory

4.4.1

Storing a reading

The STARTER 3100C can store up to 99 endpoint results.
Press
been stored.
If you press
further data you will have to clear the memory. (See 3.4.3)
4.4.2

Recalling from memory

Press and hold
measurement reaches endpoint.
Press
is being displayed.
Press
4.4.3

Clearing the memory

Pressing
Press
Now we can two choices:
Press
Press
4.5

Printing

If a printer (e.g. SF-F40A) is connected to the STARTER 3100C, a print-out is automatically
generated after each end pointed measurement or calibration. (printer baud rate 9600bps; 8 data
bit; none parity bit; 1 stop bit)
The format for the print-out following a Cond. measurement is:
End Point, Value, Temp., ATC/MTC
Auto EP, 68.8μS/cm, 26.8° C , ATC
Ref.25° C , T.Comp.Coeff. 2.00%/° C
STARTER 3100C Bench Conductivity Meter
when the measurement reaches endpoint. M01 indicates that one result has
when M99 is displayed, FUL displays to indicate the memory is full. To store
to recall the stored values from memory when the current
or
to scroll through the stored results. R01 to R99 indicates which result
to exit.
to scroll through the stored results until "
or
, CLr blinks;
to confirm the deletion of all the stored data.
to return to the measurement mode without deleting the memory.
" appears.